Chapter 5 
Analog Measurements - Screens and Control Fields
Lvl (marker)
Lvl (marker)
Lvl (marker)
This measurement field displays the signal level of the signal at the 
current marker position on the spectrum analyzer. (See 
.)
Operating Considerations
The unit-of-measure for this field is dependent on the source of the 
signal being measured. For instance, when measuring a signal from the 
AUDIO IN connector, the amplitude is measured in Volts. When 
looking at a signal from the FM demodulator, the amplitude is given in 
units of kHz. 
When the 
Vert Offset
 field is 
0.00, the displayed marker level is 
referenced to the center line generated by the vertical offset feature, not 
the center line of the screen. 
The 
Ref Set
 function can be used with this measurement to display levels 
relative to a specific value. (See 
.) 
Screens on which this field is displayed
SCOPE: all control menus
SPEC ANL: all control menus