HP Cell Phone Accessories E6380-90019 Manual Do Utilizador

Página de 228
136
Chapter 5 
Analog Measurements - Screens and Control Fields
Input Gain
Input Port
Input Gain
This field displays and selects the gain of the AF analyzer’s input 
amplifier. The choices are 0 dB, 20 dB, or 40 dB. Refer to the 
Gain 
Cntl
 field more information.
Screens on which this field is displayed
AF ANALZYER
Input Port
This field selects the RF IN/OUT or ANT IN port for making RF 
measurements. The RF IN/OUT port must be used for making TX 
power measurements. 
Operating Considerations
If the RF power at the RF IN/OUT port exceeds allowable limits, a loud 
warning signal sounds and a message appears at the top of the screen. 
If this occurs, disconnect the RF power, reset using 
Meas Reset
 (press and 
release the 
Shift
 key, then the 
Hold
 key), and allow the Test Set to cool off 
for approximately two minutes before making any other measurements 
on this port. 
The ANT IN (antenna input) connector provides a highly-sensitive 
input for very low level signals (such as “off the air” measurements). 
You cannot measure TX (RF) power on the RF ANALYZER screen 
using the ANT IN port. 
CAUTION
Connecting a signal of >60 mW to the ANT IN port can 
cause instrument damage (although internal protection 
circuits can typically withstand a short-duration signal of 
several Watts).
If the overpower circuit is triggered (signified by a 
warning message at the top of the screen), remove the 
signal from the ANT IN port, and reset using 
Meas Reset
 
(press and release the 
Shift
 key, then the 
Hold
 key) or turn 
the Test Set off and on.
Maximum signal levels at the RF IN/OUT, DUPLEX OUT, 
and ANT IN ports are printed on the connector panel. 
Exceeding these levels can cause permanent instrument 
damage.
Screens on which this field is displayed
SPEC ANL: RF IN/ANT main controls menu