Chapter 3 
Overview of the Test Set’s Built-In Tools
Clearing RAM
Clearing RAM 
RAM can be cleared using the 
RAM_MANAGER
 program in the Test Set’s 
ROM. This program clears all RAM, including any SAVE/RECALL 
registers saved to 
Internal
 RAM. 
This program is accessed by running the 
IB_UTIL
 program.
Step 1. Select the SOFTWARE MENU screen. 
Step 2. Select the 
Select Procedure Location:
 field and choose 
ROM
Step 3. Select the 
Select Procedure Filename:
 field. 
•Select 
IB_UTIL
 from the list of programs. Press 
Run Test
 to 
display a list of procedures; select 
RAM_MANAGER
 and follow 
directions to clear RAM.
To preserve the SAVE/RECALL registers, do not use the 
RAM_MANAGER
 
program. Instead, load any other test program from a card or from RAM 
to clear the previous test program, and then reload the desired test.
See the 
, and 
 for more information about 
save/recall registers.