Chapter 3
Overview of the Test Set’s Built-In Tools
Clearing RAM
Clearing RAM
RAM can be cleared using the
RAM_MANAGER
program in the Test Set’s
ROM. This program clears all RAM, including any SAVE/RECALL
registers saved to
registers saved to
Internal
RAM.
This program is accessed by running the
IB_UTIL
program.
Step 1. Select the SOFTWARE MENU screen.
Step 2. Select the
Select Procedure Location:
field and choose
ROM
.
Step 3. Select the
Select Procedure Filename:
field.
•Select
IB_UTIL
from the list of programs. Press
Run Test
to
display a list of procedures; select
RAM_MANAGER
and follow
directions to clear RAM.
To preserve the SAVE/RECALL registers, do not use the
RAM_MANAGER
program. Instead, load any other test program from a card or from RAM
to clear the previous test program, and then reload the desired test.
to clear the previous test program, and then reload the desired test.
See the
save/recall registers.