APPLICATION FUNCTIONS 
5-53
MON2000
JULY 2010
 Performing a Manual Calibration
3. To change a variable assignment, click the 
appropriate Variable cell. Use the provided 
pull-down menu and click the desired 
variable to select it. 
4. Click the 
 button to accept your 
changes and return to the Analog Outputs 
dialog.
Click the 
 button to abort and 
return to the Analog Outputs dialog.
5.10.4 Performing a Manual Calibration
To manually calibrate an analog output,
1. Select the desired analog output by clicking 
anywhere in the corresponding row.
2. Set the Zero Scale and Full Scale values as 
desired.
3. Set the Fixed/Var parameter to “Fixed”.
4. Set Fixed Value equal to the Zero Scale 
value.
5. Set Zero Adjustment and Full Adjustment to 
“0.0”.
6. Click the 
 button to accept your 
changes and exit from the Analog Outputs 
dialog.
7. Return to the Analog Outputs dialog. Wait 
until the Current Value of the analog output 
is equal to the Zero Scale value (
8. Record the value, in engineering units, read 
by the receiving device (e.g., a voltmeter).
9. Set Fixed Value equal to the Full Scale 
value (
Setting both adjustment values 
to “0.0” disables the scale 
adjustment.
 for an 
alternative calibration method 
for analog outputs. 
 includes an 
example of calibrating an 
output scale of 0 to 1200.