5-54
 APPLICATION FUNCTIONS
MON2000
Performing an Automated Calibration
JULY 2010
10.Click the 
 button to accept your 
changes and exit from the Analog Outputs 
dialog.
11.Return to the Analog Outputs dialog. Wait 
until the Current Value of the analog output 
).
12.Record the value, in engineering units, read 
by the receiving device (e.g., a voltmeter).
13.Set Zero Adjustment to the value recorded 
in 
.
14.Set Full Adjustment to the value recorded in 
15.Set the Fixed/Var parameter to “Var”.
16.Click the 
 button to complete the 
calibration and return to the main window.
5.10.5 Performing an Automated Calibration
To perform an automated analog output 
calibration, 
1. Select the desired analog output by clicking 
anywhere in the corresponding row.
2. Click the 
 button or press the 
F8
 key. 
You can abort this process at any time by 
clicking the 
 button.
3. The Zero Scale Adjustment dialog appears.
Input the appropriate value and click the 
 button.
To determine the 
uncalibrated Zero Scale and 
Full Scale analog output 
levels, see 
To prevent scale adjustment, 
set both adjustment values to 
zero (0.0).
 for additional 
instructions on calibrating 
analog outputs. 
includes a calibration example 
for an output scale of 0 to 
1200.