Emerson Process Management 3-9000-522 Manual Do Utilizador
UPGRADE 2350A GC S/W AND 2350 EPROMS
E-13
MON2000
JULY 2010
Reassembly Procedures
4. To reset the CPU registers, remove the
jumper from J14 pins 1 and 2, and place it
to short J14 pins 2 and 3.
5. After briefly shorting J14 pins 2 and 3,
remove the jumper again. Place it back in
its original position, shorting J14
pins 1 and 2.
E.7
REASSEMBLY PROCEDURES
1. Place and securely plug the SBC53 CPU
Board into slot number 2 of the Card Cage
Assembly.
2. Reconnect all ribbon cables that were
unplugged (
).
Because the ribbon cables are arranged in
layers, it important to reconnect them in the
order listed below:
Table E-1 2350 GC Ribbon Cables to Connect
Part
Number
Number
System Interface
Board connection
Board connection
Cable Originating
from Board
from Board
3-2350-063
J12
SBC53
3-2350-052
J9
SBC53
3-2350-062
J11
SBC53
3-2350-051
J4
SBC53
3-2350-064
J8
DSPI/O
3-2350-053
J10
DSPI/O
3-2350-054
J6
RTI1281
3-2350-055
J5
RTI1281