Tektronix 2200 Manual Do Utilizador

Página de 188
4-13
Servicing Precautions
Mount Terminal Block
1. Before installing the new power supply unit in the frame, mount the 
terminal block (provided in the Field Modification kit) on the rear of the 
Todd Multi-Output power assembly, as shown in 
terminal block has four terminals, but only two will be used. Remove 
and discard the four tab connectors from the terminal block, but keep 
two of the screws for the next step.
2. Install the blue and brown jumper wires from the kit as shown in 
. Connect only the power supply ends for now. (You will 
have to disconnect the wires from the terminal block again when 
connecting the frame wiring later in this procedure.
Figure 4-4.  Installation of Terminal Block and Jumpers
Using Screws
Provided,
Mount
Terminal Block
(from Kit)
on Rear of
Power Supply
Assembly
Install Blue
Jumper Wire
Install Brown
Jumper Wire
2405-01