Справочник Пользователя для Philips 862474
Part 3 Using Support Functions
6-12
Troubleshooting
Error Code
Categories
As a reference, the following table lists the general categories for other error codes. The error
code categories are shown only in hexadecimal format.
code categories are shown only in hexadecimal format.
64
The SpO
2
module is sending an
error message to the host CPU.
1. Ensure PicoSat module is
properly connected.
2. Replace the PicoSat module.
3. If problem persists, replace main
3. If problem persists, replace main
PCB module.
?
The CO
2
module is sending an
error message to the host CPU.
1. Ensure Oridion module is
properly connected.
2. Rplace the Oridion module.
3. If problem persists, replace main
3. If problem persists, replace main
PCB module.
65, 66
PicoSat SpO
2
module has detected
an error during initialization.
1. Replace PicoSat module.
2. If problem persists, replace main
2. If problem persists, replace main
PCB module.
6E-71
PicoSat SpO
2
module has detected
an error on its serial port.
1. Replace PicoSat module.
2. If problem persists, replace main
2. If problem persists, replace main
PCB module.
Hex Code
Explanation
Recommended Action
Hexadecimal Code
Explanation
500xxx
internal user interface error
501xxx
remove serial port error
502xxx
date and time error
503xxx
NBP error
504xxx
front end error
505xxx
alarm error
506xxx
audio error
507xxx
recorder error
508xxx
trend error
509xxx
flash memory data error
50axxx
SpO
2
error
50bxxx
ECG error
50cxxx
power-down task error
50dxxx
on-board diagnostic error
50exxx
power monitor error
50fxxx
temperature measurement error
510xxx
internal user interface error
511xxx
error handling error
513xxx
serial driver error