Troubleshooting
En
glis
h
LCD: Blocked Tube! 
Air path is blocked.
1. Check the air path for any blockages.
2. Remove the blockages.
3. If the alarm is not cleared, stop 
treatment. 
4. Re-start treatment. 
LCD: High Temperature [10, 11, 12, 23]!
The temperature inside the device is too 
high. Treatment may lead to stop.
Ensure the ambient temperature is within 
the specified operating range. If the 
problem persists within the specified 
operating conditions please return the 
device for servicing.
Contact your clinician.
LCD: High Pressure!
Therapy pressure exceeds pre-set alarm 
level.
1. Stop treatment. 
2. Re-start treatment. 
If the problem persists contact your 
clinician.
LCD: Low Pressure!
The air tubing is not connected properly.
1. Check the air circuit integrity and 
reconnect.
2. If the alarm doesn‘t get cleared, stop 
treatment.
3. Re-start treatment. 
LCD: Circuit disconnected!
The air circuit is not connected properly. 
1. Check the air circuit integrity and 
reconnect.
2. If the alarm is not cleared, stop 
treatment.
3. Re-start treatment. 
LCD: Low Minute Ventilation!
Minute ventilation level has dropped 
below the alarm setting level.
Contact your clinician.
LCD: Low Respiratory Rate!, High Respiratory Rate!
The respiratory rate level has dropped 
below or has exceeded the alarm setting 
level.
Contact your clinician.
LCD: High Leak!
High mask leak for more than 20 
seconds.
• Adjust the mask to minimise leak (see 
• Check the air circuit integrity and 
reconnect.
• If the problem persists contact your 
clinician.
Problem / possible cause
Action