24
LCD: Non-Vented Mask!
• Connection of a non-vented mask.
• Mask vents may be blocked.
• ResMed Leak Valve is missing or vent 
is blocked.
• Ensure the mask has vents.
• Ensure the mask vents are not 
blocked.
• Ensure the ResMed Leak Valve is 
installed and that the vent is not 
blocked.
• Ensure oxygen (if in use) has only been 
connected at the rear of the device.
• If the problem persists contact your 
clinician.
LCD: Apnoea!
The device detects an apnoea that has 
exceeded the pre-set alarm level. 
• Breathe normally to disable the alarm.
• If the problem persists contact your 
clinician.
LCD: Internal Battery Low!
The internal battery capacity is below 30%.
Connect the device to mains power.
LCD: Low SpO
2
!
SpO
2
 has dropped below pre-set alarm 
level.
• Check the attachment of the sensor.
• If the problem persists contact your 
clinician.
LCD: SpO
2
 finger sensor failure!
The finger sensor is not connected 
properly or delivers faulty values.
Check if the finger sensor is attached 
properly to the finger and connection to 
the pulse oximeter.
LCD: Xpod oximeter disconnected!
The pulse oximeter is disconnected. 
Check if the pulse oximeter is connected 
properly to the device.
LCD: Low FiO
2
 Level!
FiO
2
 has dropped below the pre-set 
alarm level. 
• Perform FiO
2
 sensor calibration.
• If the problem persists contact your 
clinician.
LCD: High FiO
2
 Level!  
FiO
2
 has exceeded the pre-set alarm 
level. 
• Perform FiO
2
 sensor calibration.
• If the problem persists contact your 
clinician. 
LCD: Keypad Failure! 
One of the keys was held down for more 
than 10 seconds or got stuck. 
Remove any blockages from the keypad.
LCD: Attention High Temp [42, 43, 44, 45]!
The temperature inside the device is 
high. 
Ensure the ambient temperature 
conditions are within the specified 
operating range. 
LCD: Internal Battery Use!
The device is using the internal battery. 
Check if the power cord is properly 
connected to the device if you want to 
run from mains power.
Press the Alarm mute button 
 to cancel 
the alarm.
Problem / possible cause
Action