Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
98 
Example: If all passwords are enabled, the following keypad sequence would be required to enter the 
SETUP menu: 
 
SAMPLE        RANGE = 500.000 PPM      CO2 =X.XXX  
 
 < TST  TST >  CAL
          
SETUP
 
 
 
 SAMPLE  
ENTER SETUP PASS: 0
 
  0         0          0
 
 
     
      
   ENTR     EXIT
 
prompts for password 
number 
See Table 6-8 for 
Passwords and Levels 
Example: this 
password enables  the 
SETUP mode
 
 
 
 SAMPLE  
ENTER SETUP P ASS: 0
 
 
  8         1          8
 
 
     
    
    ENTR    EXIT
 
 SETUP X.X      
PRIMARY SETUP MENU
 
 
 CFG  DAS  RNGE  PASS  CLK  MORE 
 
EXIT
 
Press individual 
buttons to set 
numbers 
 
Note that the instrument still prompts for a password when entering the VARS and DIAG menus, even if 
passwords are disabled, but it displays the default password (818) upon entering these menus.  The user 
only has to press ENTR to access the password-protected menus but does not have to enter the required 
number code.