Справочник Пользователя для Teledyne gfc 7000ta
Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
100
In order to compensate for CPU clocks which run faster or slower, you can adjust a variable called
CLOCK_ADJ
CLOCK_ADJ
to speed up or slow down the clock by a fixed amount every day. To change this variable,
press:
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X 0 ) DAS_HOLD_OFF=15.0 Minutes
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SAMPLE RANGE = 500.000 PPM
CO2
=X.XXX
< TST TST > CAL
SETUP
3x EXIT returns
to the main SAMPLE display
Enter sign and number of seconds per
day the clock gains (-) or loses (+).
SETUPX.X 1 ) CONC_PRECISION = 3
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X
4) CLOCK_ADJ=0 Sec/Day
PREV JUMP
EDIT
PRNT EXIT
SETUP X.X
CLOCK_ADJ:
0 Sec/Day
+ 0 0
ENTR
EXIT
SETUP X.X
4) CLOCK_ADJ=0 Sec/Day
PREV NEXT JUMP
EDIT PRNT
EXIT
Continue to press NEXT until …