Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
114 
6.11.7. Multidrop RS-232 Set Up 
The RS-232 multidrop consists of a printed circuit assembly that is seated on the analyzer’s CPU card 
(see Figures 6-10 and 6-13) and is connected to the CPU and the motherboard.  This PCA includes all 
circuitry required to enable your analyzer for multidrop operation.  It converts the instrument’s RS232 port 
to multidrop configuration allowing up to eight analyzers to be connected the same I/O port of the host 
computer.   
Because both of the DB9 connectors on the analyzer’s back panel are needed to construct the multidrop 
chain, COM2 is no longer available for separate RS-232 or RS-485 operation; however, the USB com 
port is available. 
 
 
Figure 6-9: 
Multidrop/LVDS PCA Seated on CPU 
 
Each analyzer in the multidrop chain must have: 
  One Teledyne Instruments option 62 installed.   
  One 6’ straight-through, DB9 male  DB9 Female cable (Teledyne Instruments P/N WR0000101) is 
required for each analyzer. 
To set up the network, for each analyzer: 
1.  With NO power to the instrument, open the rear panel and locate the CPU. 
2.  On the LVDS card, disconnect the cable that’s connected to J1 and remove the LVDS card from 
the CPU card. 
3.  Mount the Multidrop/LVDS card onto the CPU and reconnect the LVDS cable to J1.