Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
115 
4.  Connect a cable between the Multidrop/LVDS card J4 and the motherboard P12. 
5.  Connect a cable between the Multidrop/LVDS card J3 and the CPU COM1 connector. 
6.  Close the instrument. 
7.  Using straight-through, DB9 male  DB9 Female cable, interconnect the host and the analyzers 
as shown in Figure 6-14. 
8.  BEFORE communicating from the host, power on the instruments and check that the Machine ID 
code is unique for each (see Section Analyzer ID). On the front panel menu, use 
SETUP>MORE>COMM>ID. Note that the default ID is typically the model number; to change the 
4-digit identification number, press the button of the digit to be changed).. 
NOTE: 
 Teledyne Instruments recommends setting up the first link, between the Host and 
the first analyzer and testing it before setting up the rest of the chain.   
 
 
 
 
Analyzer 
Analyzer 
Analyzer 
Last Analyzer 
Female DB9
Male DB9 
RS-232 
COM2 
RS-232 
COM2 
RS-232
COM2
RS-232
COM2
 Host 
 
RS-232 port 
Make Sure 
Jumper between 
JP2 pins 21 
 22
is installed. 
KEY: 
 
Figure 6-10: 
RS232-Multidrop PCA Host/Analyzer Interconnect Diagram