Справочник Пользователя для Teledyne gfc 7000ta
Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
118
6.12. SETUP
MORE VARS: Internal Variables (VARS)
The GFC 7000TA has several-user adjustable software variables, which define certain operational
parameters. Usually, these variables are automatically set by the instrument’s firmware, but can be
manually re-defined using the VARS menu. Table 6-14 lists all variables that are available within the 818
password protected level.
parameters. Usually, these variables are automatically set by the instrument’s firmware, but can be
manually re-defined using the VARS menu. Table 6-14 lists all variables that are available within the 818
password protected level.
Table 6-14: Variable Names (VARS) Revision B.3
NO. VARIABLE
DESCRIPTION
ALLOWED
VALUES
0 DAS_HOLD_OFF
Changes the internal data acquisition system (DAS) hold-off
time, which is the duration when data are not stored in the
DAS because the software considers the data to be
questionable. That is the case during warm-up or just after
the instrument returns from one of its calibration modes to
SAMPLE mode. DAS_HOLD_OFF can be disabled entirely
in each DAS channel.
time, which is the duration when data are not stored in the
DAS because the software considers the data to be
questionable. That is the case during warm-up or just after
the instrument returns from one of its calibration modes to
SAMPLE mode. DAS_HOLD_OFF can be disabled entirely
in each DAS channel.
Can be between 0.5
and 20 minutes
Default=15 min.
1 CONC_PRECISION
Allows the user to set the number of significant digits to the
right of the decimal point display of concentration and
stability values.
right of the decimal point display of concentration and
stability values.
AUTO, 1, 2, 3, 4
Default=AUTO
2
STABIL_GAS
Selects which gas measurement is displayed when the STABIL
test function is selected.
test function is selected.
CO
2
& O
2
1
3 DYN_ZERO
Dynamic zero automatically adjusts offset and slope of the
CO
CO
2
response when performing a zero point calibration
during an AutoCal (Section 7).
ON/OFF
4 DYN_SPAN
Dynamic span automatically adjusts slope and slope of the
CO
CO
2
response when performing a zero point calibration
during an AutoCal (Section 7).
Note that the DYN_ZERO and DYN_SPAN features are not
allowed for applications requiring EPA equivalency.
allowed for applications requiring EPA equivalency.
ON/OFF
5 CLOCK_ADJ
Adjusts the speed of the analyzer’s clock. Choose the +
sign if the clock is too slow, choose the - sign if the clock is
too fast.
sign if the clock is too slow, choose the - sign if the clock is
too fast.
-60 to +60 s/day
1
O
2
option is only available in analyzers with O
2
sensor options installed.