Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
118 
6.12. SETUP 
 MORE  VARS: Internal Variables (VARS) 
The GFC 7000TA has several-user adjustable software variables, which define certain operational 
parameters.  Usually, these variables are automatically set by the instrument’s firmware, but can be 
manually re-defined using the VARS menu.  Table 6-14 lists all variables that are available within the 818 
password protected level. 
 
Table 6-14:   Variable Names (VARS) Revision B.3 
NO. VARIABLE 
DESCRIPTION 
ALLOWED 
VALUES 
0 DAS_HOLD_OFF 
Changes the internal data acquisition system (DAS) hold-off 
time, which is the duration when data are not stored in the 
DAS because the software considers the data to be 
questionable.  That is the case during warm-up or just after 
the instrument returns from one of its calibration modes to 
SAMPLE mode.  DAS_HOLD_OFF can be disabled entirely 
in each DAS channel. 
Can be between 0.5 
and 20 minutes 
Default=15 min. 
1 CONC_PRECISION 
Allows the user to set the number of significant digits to the 
right of the decimal point display of concentration and 
stability values. 
AUTO, 1, 2, 3, 4 
Default=AUTO 
2
 
STABIL_GAS 
Selects which gas measurement is displayed when the STABIL 
test function is selected. 
CO
2
 & O
2
3 DYN_ZERO 
Dynamic zero automatically adjusts offset and slope of the 
CO
2
 response when performing a zero point calibration 
during an AutoCal (Section 7). 
ON/OFF 
4 DYN_SPAN 
Dynamic span automatically adjusts slope and slope of the 
CO
2
 response when performing a zero point calibration 
during an AutoCal (Section 7). 
Note that the DYN_ZERO and DYN_SPAN features are not 
allowed for applications requiring EPA equivalency. 
ON/OFF 
5 CLOCK_ADJ 
Adjusts the speed of the analyzer’s clock.  Choose the + 
sign if the clock is too slow, choose the - sign if the clock is 
too fast. 
-60 to +60 s/day 
1
 O
2
 option is only available in analyzers with O
2
 sensor options installed.