6.13. SETUP 
  MORE  DIAG: Using the Diagnostics 
Functions 
A series of diagnostic tools is grouped together under the SETUPMOREDIAG menu.  As these 
parameters are dependent on firmware revision (see Menu Tree A-6 in Appendix A).  The individual 
parameters, however, are explained in more detail in the indicated in 6-15.  These tools can be used in a 
variety of troubleshooting and diagnostic procedures and are referred to in many places of the 
maintenance and trouble-shooting s. 
Table 6-15:   GFC 7000TA Diagnostic (DIAG) Functions 
DIAGNOSTIC FUNCTION AND MEANING 
Front Panel 
Mode Indicator 
 
SIGNAL I/O
: Allows observation of all digital and analog signals in 
the instrument.  Allows certain digital signals such as valves and 
heaters to be toggled ON and OFF. 
DIAG I/O 
6.13.2 
ANALOG OUTPUT
: When entered, the analyzer performs an analog 
output step test.  This can be used to calibrate a chart recorder or 
to test the analog output accuracy. 
DIAG AOUT 
6.13.3 
ANALOG I/O CONFIGURATION
: the signal levels of the instruments 
analog outputs may be calibrated (either individually or as a 
group).  Various electronic parameters such as signal span, and 
offset are available for viewing and configuration. 
DIAG AIO 
6.13.4  
ELECTRIC TES
T: The analyzer is performing an electric test.  This 
test simulates IR detector signal in a known manner so that the 
proper functioning of the sync/demod board can be verified. 
DIAG OPTIC 
6.13.5 
DARK CALIBRATION
:  The analyzer is performing a dark 
calibration procedure.  This procedure measures and stores the 
inherent dc offset of the sync/demod board electronics. 
DIAG ELEC 
6.13.6 
PRESSURE CALIBRATION
: The analyzer records the current 
output of the sample gas pressure sensor.  This value is used by 
the CPU to compensate the CO
2
 concentration. 
DIAG PCAL 
6.13.7 
FLOW CALIBRATION
: This function is used to calibrate the gas flow 
output signal.  This setting is retained when exiting DIAG
DIAG FCAL 
6.13.8 
TEST CHAN OUTPUT
: Configures the A4 analog output channel. 
DIAG TCHN 
6.13.9