Справочник Пользователя для Teledyne gfc 7000ta
Operating Instructions
Model GFC7000TA Carbon Dioxide Analyzer
Teledyne Analytical Instruments
120
6.13. SETUP
MORE DIAG: Using the Diagnostics
Functions
A series of diagnostic tools is grouped together under the SETUPMOREDIAG menu. As these
parameters are dependent on firmware revision (see Menu Tree A-6 in Appendix A). The individual
parameters, however, are explained in more detail in the indicated in 6-15. These tools can be used in a
variety of troubleshooting and diagnostic procedures and are referred to in many places of the
maintenance and trouble-shooting s.
parameters are dependent on firmware revision (see Menu Tree A-6 in Appendix A). The individual
parameters, however, are explained in more detail in the indicated in 6-15. These tools can be used in a
variety of troubleshooting and diagnostic procedures and are referred to in many places of the
maintenance and trouble-shooting s.
Table 6-15: GFC 7000TA Diagnostic (DIAG) Functions
DIAGNOSTIC FUNCTION AND MEANING
Front Panel
Mode Indicator
SIGNAL I/O
: Allows observation of all digital and analog signals in
the instrument. Allows certain digital signals such as valves and
heaters to be toggled ON and OFF.
heaters to be toggled ON and OFF.
DIAG I/O
6.13.2
ANALOG OUTPUT
: When entered, the analyzer performs an analog
output step test. This can be used to calibrate a chart recorder or
to test the analog output accuracy.
to test the analog output accuracy.
DIAG AOUT
6.13.3
ANALOG I/O CONFIGURATION
: the signal levels of the instruments
analog outputs may be calibrated (either individually or as a
group). Various electronic parameters such as signal span, and
offset are available for viewing and configuration.
group). Various electronic parameters such as signal span, and
offset are available for viewing and configuration.
DIAG AIO
6.13.4
ELECTRIC TES
T: The analyzer is performing an electric test. This
test simulates IR detector signal in a known manner so that the
proper functioning of the sync/demod board can be verified.
proper functioning of the sync/demod board can be verified.
DIAG OPTIC
6.13.5
DARK CALIBRATION
: The analyzer is performing a dark
calibration procedure. This procedure measures and stores the
inherent dc offset of the sync/demod board electronics.
inherent dc offset of the sync/demod board electronics.
DIAG ELEC
6.13.6
PRESSURE CALIBRATION
: The analyzer records the current
output of the sample gas pressure sensor. This value is used by
the CPU to compensate the CO
the CPU to compensate the CO
2
concentration.
DIAG PCAL
6.13.7
FLOW CALIBRATION
: This function is used to calibrate the gas flow
output signal. This setting is retained when exiting DIAG.
DIAG FCAL
6.13.8
TEST CHAN OUTPUT
: Configures the A4 analog output channel.
DIAG TCHN
6.13.9