Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
138 
6.13.7. Pressure Calibration 
A sensor at the exit of the sample chamber continuously measures the pressure of the sample gas.  This 
data is used to compensate the final CO
2
 concentration calculation for changes in atmospheric pressure 
and is stored in the CPU’s memory as the test function PRES (also viewable via the front panel).   
NOTE 
This calibration must be performed when the pressure of the sample gas is equal 
to ambient atmospheric pressure.   
 Before performing the following pressure calibration procedure, disconnect the 
sample gas pump and the sample gas-line vent from the sample gas inlet on the 
instrument’s rear panel. 
To cause the analyzer to measure and record a value for PRES, press. 
 
 
 
 
 
SAMPLE          RANGE = 500.000 PPM        CO2 =X.XXX  
 
 < TST  TST >  CAL  
 
 
SETUP
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  
RNGE
  
PASS  CLK  
MORE
  
   EXIT
 
SETUP X.X     
SECONDARY SETUP MENU
  
 
 
 
COMM  VARS  
DIAG
   ALRM       
                   EXIT
 
Adjust these values until the 
displayed pressure equals the 
pressure measured by the 
independent pressure meter. 
 
 
 
DIAG  
SIGNAL I / O
 
NEXT 
      
          
ENTR 
  
EXIT
 
Exit at 
any time 
to return 
to main 
the 
SETUP 
menu 
SETUP X.X  
      ENTER DIAG PASS: 818 
 
    8     1     8 
      
        
    ENTR  
EXIT 
Repeat Pressing NEXT until . .
 .
 
DIAG PCAL     
ACTUAL PRESS : 27.20 IN-HG-A
 
     2       7       .2       0                           ENTR  EXIT
 
ENTR accepts the 
new value and 
returns to the 
previous menu 
EXIT 
ignores the 
new value and 
returns to the 
previous menu