Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
140 
6.13.9. Test Channel Output 
When activated, output channel A4 can be used to report one of the test functions viewable from the 
SAMPLE mode display.  To activate the A4 channel and select a test function, follow this button 
sequence: 
SETUP X.X  
      ENTER DIAG PASS: 818 
 
    8     1     8 
      
        
    ENTR  
EXIT 
SAMPLE*          RANGE = 500.000 PPM       CO2 =X.XXX   
 
 < TST  TST >  CAL  
 
 
SETUP
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  
RNGE
  
PASS  CLK  
MORE
  
   EXIT
 
 
 
 
DIAG  
 
 
 
SIGNAL I / O  
 
            NEXT 
      
                        
ENTR 
 
EXIT
 
DIAG  
ANALOG OUTPUT  
 
PREV
  NEXT 
      
                        
ENTR
  
EXIT
 
 
 
 
DIAG TCHN 
 
TEST CHANNEL: NONE 
 
 
NEXT 
      
          ENTR
   
   
EXIT
 
 
 
 
DIAG 
              
TEST CHAN OUTPUT  
 
PREV
 
NEXT
 
      
          ENTR
 
  
   
EXIT
 
Continue to press NEXT until … 
 
 
 
DIAG TCHN 
TEST CHANNEL: CO2 MEASURE 
 
PREV 
NEXT 
      
         
 ENTR   
   EXIT
 
Press PREV or NEXT 
to move through the 
list of available 
parameters 
(Table 6-13) 
Press EXIT to 
return to the 
DIAG
 menu 
Press ENTR to select 
the displayed 
parameter activating 
the test channel. 
SETUP X.X     
SECONDARY SETUP MENU
  
 
 
 
COMM  VARS  
DIAG
   
 ALRM
 
      
   EXIT
 
 
 
Table 6-22:   Test Parameters Available for Analog Output A4 
TEST CHANNEL 
ZERO 
FULL SCALE 
NONE 
Test Channel is turned off 
CO
2
 MEASURE 
0 mV 
5000 mV* 
CO
2
 REFERENCE 
0 mV 
5000 mV* 
SAMPLE PRESS 
0 "Hg 
40 "Hg 
SAMPLE FLOW 
0 cc/m 
1000 cc/m 
SAMPLE TEMP 
0
C 70C 
BENCH TEMP 
0
C 70C 
WHEEL TEMP 
0
C 70C 
CHASSIS TEMP 
0
C 70C 
PHT DRIVE 
0 mV 
5000 mV