Calibration Procedures 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
165 
7.3. Manual Calibration Checks 
Informal calibration checks, which only evaluate but do not alter the analyzer’s response curve, are 
recommended as a regular maintenance item and in order to monitor the analyzer’s performance.  To 
carry out a calibration check rather than a full calibration, follow these steps. 
STEP ONE
: Connect the sources of zero air and span gas as shown in Figures 7.1 or 7.2. 
STEP TWO
: Perform the zero/span calibration check procedure: 
 
SAMPLE               
STABIL=XXX.X PPM
     
   CO2=X.XXX 
 
 
 < TST  TST > 
 CAL 
 
 
 
SETUP
 
ACTION: 
Record the CO
2
 
concentration 
reading. 
ACTION:
Supply span gas to the instrument 
SAMPLE              
STABIL=XXX.X PPM
      
   CO2=X.XXX 
 
 < TST  TST > 
 CAL 
 
 
 
SETUP
 
ACTION: 
Record the CO
2
 
concentration 
reading. 
Wait until 
STABIL
 is 
below 1.0 ppm.  
This may take 
several minutes. 
The value of 
STABIL
 may jump 
significantly. 
Wait until it falls 
below 1.0 ppm.  This 
may take several 
minutes. 
SAMPLE              
STABIL=XXX.X PPM
          CO2=X.XXX 
 < TST  TST >
 
 CAL 
 
 
 
SETUP
 
The SPAN key appears during the transition from zero to 
span. You may see both SPAN and ZERO buttons. 
SAMPLE               RANGE = 500.0 PPM 
       CO2=X.XXX
 < TST  TST >
 
 CAL 
 
 
 
SETUP
 
Scroll the display to the 
STABIL
 test  function. 
ACTION:
Supply  the instrument with zero  gas.