Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Calibration Procedures 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
167 
Source of
SAMPLE GAS
Removed during 
calibration
VENT
Calibrated
 CO
2
Gas
at span gas 
concentration
VENT 
here if input 
is pressurized
Chassis
PRESSURE SPAN
IZS
PURGE LINE
SAMPLE
EXHAUST
VENT SPAN
External 
Zero Air 
Scrubber
N
2
 
Figure 7-4: 
Pneumatic Connections – Ambient Zero/Pressurized Span Valves and External 
Zero Air Scrubber 
 
Figure 7-5: 
Pneumatic Connections – Ambient Zero/Ambient Span Valves