Maintenance 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
Teledyne Analytical Instruments 
181 
Table 8-1:     GFC 7000TA Maintenance Schedule 
ITEM ACTION  FREQ 
CAL 
CHECK 
REQ’D. 
MANUAL  
DATE PERFORMED 
 
 
 
 
 
 
 
 
 
 
Particulate 
Filter 
Replace 
Weekly or as 
needed 
Yes 
 
 
 
 
 
 
 
 
 
 
 
Verify Test 
Functions 
Record and 
analyze 
Weekly or after 
any 
Maintenance 
or Repair  
No 
 
 
 
 
 
 
 
 
 
 
 
Pump 
Diaphragm 
Replace Annually  Yes 
 
 
 
 
 
 
 
 
 
 
 
Perform 
Flow Check 
Check Flow 
Annually  
No 
 
 
 
 
 
 
 
 
 
 
 
Perform 
Leak Check 
Verify Leak 
Tight 
Annually or 
after any 
Maintenance 
or Repair 
No 
 
 
 
 
 
 
 
 
 
 
 
Pneumatic 
lines 
Examine 
and clean 
As needed 
Yes if 
cleaned 
 
 
 
 
 
 
 
 
 
 
 
Cleaning Clean As 
needed 
Only if 
cover 
remv’d