Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Maintenance 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
Teledyne Analytical Instruments 
183 
8.2. Predicting Failures Using the Test Functions 
The Test Functions can be used to predict failures by looking at how their values change over time.  
Initially it may be useful to compare the state of these Test Functions to the values recorded on the 
printed record of the final calibration performed on your instrument at the factory, p/n 04307.  Table 9-3 
can be used as a basis for taking action as these values change with time.  The internal data acquisition 
system (DAS) is a convenient way to record and track these changes.  Use APICOM to download and 
review this data from a remote location. 
  
Table 8-3:     Predictive uses for Test Functions 
FUNCTION CONDITION 
BEHAVIOR 
INTERPRETATION 
STABILITY 
Zero Cal 
Increasing 
Pneumatic Leaks – instrument & sample system 
Detector deteriorating 
CO2 MEAS 
Zero Cal 
Decreasing 
Source Aging 
Detector deteriorating 
Optics getting dirty or contaminated 
MR RATIO 
Zero Cal 
Increasing 
Source Aging 
Detector deteriorating 
Contaminated zero gas (H2O) 
Decreasing 
Source Aging 
Detector deteriorating 
GFC Wheel Leaking  
Pneumatic Leaks 
Contaminated zero gas (CO
2
Span Cal 
Increasing 
Source Aging 
Pneumatic Leaks – instrument & sample system 
Calibration system deteriorating 
Decreasing 
Source Aging 
GFC Wheel Leaking 
Calibration system deteriorating 
PRES 
Sample 
Increasing > 1” 
Pneumatic Leak between sample inlet and Sample Cell 
Change in sampling manifold 
Decreasing > 1” 
Dirty particulate filter 
Pneumatic obstruction between sample inlet and Sample 
Cell 
Obstruction in sampling manifold 
PHT DRIVE 
Any, but with 
Bench Temp at 
48°C 
Increasing 
Mechanical Connection between IR-Detector and Sample 
Cell deteriorating 
IR-Photodetector deteriorating 
OFFSET 
Zero Cal 
Increasing 
See MR Ratio - Zero Cal Decreasing above 
Decreasing 
See MR Ratio - Zero Cal Increasing above 
SLOPE 
Span Cal 
Increasing 
See MR Ratio - Span Cal Decreasing above 
Decreasing 
See MR Ratio – Span Cal Increasing above