Theory of Operation 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
193 
9.2. Pneumatic Operation 
 
CAUTION 
It is important that the sample airflow system is both leak tight and not pressurized 
over ambient pressure. 
Regular leak checks should be performed on the analyzer as described in the 
maintenance schedule, Table 9-1. 
Procedures for correctly performing leak checks can be found in Section 9.3.3. 
 
An internal pump evacuates the sample chamber creating a small vacuum that draws sample gas into the 
analyzer.  Normally the analyzer is operated with its inlet near ambient pressure either because the 
sample is directly drawn at the inlet or a small vent is installed at the inlet.  There are several advantages 
to this “pull through” configuration. 
  By placing the pump down stream from the sample chamber several problems are avoided.   
  First the pumping process heats and compresses the sample air complicating the measurement 
process.   
  Additionally, certain physical parts of the pump itself are made of materials that might chemically 
react with the sample gas.   
  Finally, in certain applications where the concentration of the target gas might be high enough to be 
hazardous, maintaining a negative gas pressure relative to ambient means that should a minor leak 
occur, no sample gas will be pumped into the atmosphere surrounding analyzer.