Theory of Operation 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
195 
pressure is greater than 2:1, the velocity of the gas through the orifice reaches the speed of sound.  As 
long as that ratio stays at least 2:1 the gas flow rate is unaffected by any fluctuations, surges, or changes 
in downstream pressure because such variations only travel at the speed of sound themselves and are 
therefore cancelled out by the sonic shockwave at the downstream exit of the critical flow orifice.   
 
SPRING
O-RINGS
FILTER
CRITICAL 
FLOW 
ORIFICE 
AREA OF 
LOW 
PRESSURE
AREA OF 
HIGH 
PRESSURE
Sonic 
Shockwave
 
Figure 9-8: 
Flow Control Assembly & Critical Flow Orifice 
The actual flow rate of gas through the orifice (volume of gas per unit of time), depends on the size and 
shape of the aperture in the orifice.  The larger the hole, the more gas molecules, moving at the speed of 
sound, pass through the orifice.  Because the flow rate of gas through the orifice is only related to the 
minimum 2:1 pressure differential and not absolute pressure the flow rate of the gas is also unaffected by 
degradations in pump efficiency due to age.   
The critical flow orifice used in the Model GFC 7000TA is designed to provide a flow rate of 800 cm