Figure 5-1: 
Current Loop Option Installed on the Motherboard 
5.2.1. Converting Current Loop Analog Outputs to Standard Voltage 
Outputs 
NOTE 
Servicing or handling of circuit components requires electrostatic discharge 
protection, i.e.  ESD grounding straps, mats and containers.  Failure to use ESD 
protection when working with electronic assemblies will void the instrument 
warranty.   
See Section 12 for more information on preventing ESD damage. 
 
To convert an output configured for current loop operation to the standard 0 to 5 VDC output operation: 
1.  Turn off power to the analyzer.   
2.  If a recording device was connected to the output being modified, disconnect it. 
3.  Remove the top cover