 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
53 
  Remove the set screw located in the top, center of the rear panel 
  Remove the screws fastening the top cover to the unit (four per side). 
  Slide the cover back and lift straight up . 
4.  Disconnect the current loop option PCA from the appropriate connector on the motherboard (see 
Figure 5-1). 
5.  Place a shunt between the leftmost two pins of the connector (see Figure 5-1). 
6.  Reattach the top case to the analyzer. 
7.  The analyzer is now ready to have a voltage-sensing, recording device attached to that output 
 
 
5.3. Expendable Kits (Options 42C, 42D and 43) 
OPTION NUMBER 
DESCRIPTION 
OPT 42C 
1 year’s supply of replacement of 47mm dia.  particulate filters 
OPT 42D 
1 full replacement’s volume of indicating soda-lime for the external CO
2
 scrubber 
included with options 51 & 53 (approximate active lifetime: 1 year) 
OPT 43 
Options 42 C & 42D