MODEL 5081-P pH/ORP
SECTION 9.0
CALIBRATION OF ORP MEASUREMENTS
SECTION 9.0
CALIBRATION OF ORP MEASUREMENTS
9.1
General
9.2
Entering and Leaving the Calibrate Menu
9.3
Using the Hold Function
9.4
Temperature Calibration
9.5
Standardization
9.1  GENERAL
The Calibrate menu allows the user to calibrate the ORP and temperature response of the sensor. 
The ORP calibration is a one-point standardization against an ORP standard. The temperature calibration is a
one-point standardization against a reference thermometer. Prompts guide the user through the calibration proce-
dures.
9.2.  ENTERING AND LEAvING THE CALIBRATE MENU
Press  CAL  on the infrared remote controller (IRC) to enter the Calibrate menu. To store new settings in memory,
press  ENTER . To leave the Calibrate menu without storing new values, press EXIT.  Pressing EXIT with a prompt
showing returns the display to the first prompt in the sub-menu. Pressing EXIT a second time returns the trans-
mitter to the process display.
If program settings are protected with a security code, pressing PROG or CAL will cause the Id screen to appear.
Key in the security code and press ENTER. The first sub-menu will appear. For more information, see Section 5.8,
Security.
A transmitter adjacent to the one being calibrated may pick up signals from the IRC. To avoid accidentally chang-
ing settings, use a different security code for each nearby transmitter. 
9.3   USING THE HOLD FUNCTION
during  calibration,  the  sensor  may  be  exposed  to  solutions  having  an  ORP  outside  the  normal  range  of  the
process. To prevent false alarms and possible undesired operation of chemical dosing pumps, place the transmit-
ter in hold during calibration. Activating HOLd keeps the transmitter output at the last value or sends the output to
a previously determined value.  
After calibration, reinstall the sensor in the process stream. Wait until readings have stabilized before deactivating
Hold.
To activate or deactivate Hold, do the following:
1. Press HOLd on the IRC.
2. The HoLd prompt appears in the display. Press é or  ê to toggle the Hold function between On and OFF.
3. Press ENTER to save the setting.