Справочник Пользователя для Emerson 5081-P

Скачать
Страница из 156
PROGRAM
tESt
EXIT
ENTER
1
2
.00
PROGRAM
dPn
EXIT
ENTER
0
00
MODEL 5081-P PH/ORP
SECTION 8.0
PROGRAMMING FOR pH MEASUREMENTS
8.9 GENERATING A TEST CURRENT (for 5081-P-HT only)
8.9.1 Purpose
This section describes how to generate output currents for testing recorders and data han-
dling systems.
8.9.2 what happens while the transmitter is generating a test current?
1. The output current goes to the programmed test value and remains there until the 
TEST function is disabled.
2. The main display continues to show the pH  of the process stream. The word HOLd
appears in the display.
3. The test current value supersedes both the HOLd value and the FAuLT value.
4. If a fault occurs while the transmitter is generating the test current, the word fault 
appears in the display and the display flashes.
8.9.3 Procedure
1. Press  PROG  on  the  infrared  remote  controller  (IRC).  OUtPUt will  appear.  Press
ENTER
2. Press NEXT until the dPn sub-menu appears. Press ENTER .
3. The tESt prompt appears. use the editing keys to change the number to the desired
value. The allowed values are between 3.80 mA and 22.00 mA. 
4. Press ENTER to start the test current. HOLD will appear o the left side of the screen.
5. To end the test current, press EXIT.
6. Press EXIT to return to the process display.
91