NI 5441 Specifications
Delay from 
Start Trigger to 
CH 0 Analog 
Output with 
OSP Disabled.
DAC Interpolation Factor
Typical Delay
Refer to the t
s2
 
documentation in 
the NI Signal 
Generators Help
 
by navigating to 
NI Signal 
Generators 
Help»Devices»
NI 5441»
Triggering»
Trigger Timing
.
Digital Interpolation Filter 
disabled.
44 Sample Clock 
Periods + 110 ns
2
58 Sample Clock 
Periods + 110 ns
4
64 Sample Clock 
Periods + 110 ns
8
65 Sample Clock 
Periods + 110 ns
Delay from 
Start Trigger to 
Digital Data 
Output with 
OSP Disabled.
40 Sample Clock periods + 110 ns.
Additional 
Delay for 
Function 
Generator 
Mode.
Add 33 Sample Clock Periods
(Applicable to Delay from Start Trigger to CH0 Analog 
Output and Delay from Start Trigger to Digital Data 
Output.)
Additional 
Delay with 
OSP Enabled.
Add 70 Sample Clock Periods for Real data processing 
mode
Add 73 Sample Clock Periods for Complex data processing 
mode.
(Applicable to Delay from Start Trigger to CH0 Analog 
Output and Delay from Start Trigger to Digital Data 
Output.)
FIR and CIC 
filters enabled.
Trigger Exporting
Exported 
Trigger 
Destinations
A signal used as a trigger can be routed out to any 
destination listed in the 
 specification of the 
Exported 
Trigger Delay
65 ns (typical). Refer to the t
s3
 documentation in the 
NI Signal Generators Help by navigating to NI Signal 
Generators Help»Devices»NI 5441»Triggering»Trigger 
Timing
.
Exported 
Trigger Pulse 
Width
>150 ns. Refer to the t
s4
 documentation in the NI Signal 
Generators Help by navigating to NI Signal Generators 
Help»Devices»NI 5441»Triggering»Trigger Timing
.
Specification