Справочник Пользователя для National Instruments 5441

Скачать
Страница из 44
NI 5441 Specifications
24
ni.com
Markers
Specification
Value
Comments
Destinations
1. PFI<0..1> (SMB front panel connectors)
2. PFI<4..5> (DIGITAL DATA & CONTROL front panel 
connector)
3. PXI_Trig<0..6> (backplane connector)
Quantity
One Marker per Segment.
Quantum
Marker position must be placed at an integer multiple of 
four samples (two samples for Complex (IQ) data).
Width
>150 ns. Refer to the t
m2
 documentation in the NI Signal 
Generators Help by navigating to NI Signal Generators 
Help»Devices»NI 5441»Waveform Generation»Marker 
Events
.
Skew
Destination
With Respect to 
Analog Output
With Respect to 
Digital Data 
Output
Refer to the t
m1
 
documentation in 
the NI Signal 
Generators Help
 
by navigating to 
NI Signal 
Generators 
Help»Devices»
NI 5441»
Waveform 
Generation»
Marker Events
.
PFI<0..1>
±2 Sample Clock 
Periods
N/A
PFI<4..5>
N/A
<2 ns
PXI_Trig<0..6>
±2 Sample Clock 
Periods
N/A
Jitter
20 ps rms