Справочник Пользователя для National Instruments 5441
NI 5441 Specifications
24
ni.com
Markers
Specification
Value
Comments
Destinations
1. PFI<0..1> (SMB front panel connectors)
2. PFI<4..5> (DIGITAL DATA & CONTROL front panel
connector)
3. PXI_Trig<0..6> (backplane connector)
—
Quantity
One Marker per Segment.
—
Quantum
Marker position must be placed at an integer multiple of
four samples (two samples for Complex (IQ) data).
four samples (two samples for Complex (IQ) data).
—
Width
>150 ns. Refer to the t
m2
documentation in the NI Signal
Generators Help by navigating to NI Signal Generators
Help»Devices»NI 5441»Waveform Generation»Marker
Events.
Help»Devices»NI 5441»Waveform Generation»Marker
Events.
—
Skew
Destination
With Respect to
Analog Output
With Respect to
Digital Data
Output
Refer to the t
m1
documentation in
the NI Signal
Generators Help
by navigating to
NI Signal
Generators
Help»Devices»
NI 5441»
Waveform
Generation»
Marker Events.
the NI Signal
Generators Help
by navigating to
NI Signal
Generators
Help»Devices»
NI 5441»
Waveform
Generation»
Marker Events.
PFI<0..1>
±2 Sample Clock
Periods
N/A
PFI<4..5>
N/A
<2 ns
PXI_Trig<0..6>
±2 Sample Clock
Periods
N/A
Jitter
20 ps rms
—