Справочник Пользователя для National Instruments NI PXIe-1065

Скачать
Страница из 75
Chapter 2
Installation and Configuration
© National Instruments Corporation
2-5
rated ambient temperature as stated in Appendix A, 
. If the 
temperature exceeds the stated spec the power switch LED will blink green, 
as discussed in the 
 section of this 
chapter.
Setting Fan Speed
The fan-speed selector switch is on the rear panel of the NI PXIe-1065 
chassis. Refer to Figure 1-2, 
locate the fan-speed selector switch. Select High for maximum cooling 
performance or Auto for improved acoustic performance. When set to 
Auto, the fan speed is determined by chassis intake air temperature.
Installing Filler Panels
To maintain proper module cooling performance, install filler panels 
(provided with the chassis) in unused or empty slots. Secure with the 
captive mounting screws provided.
Installing Slot Blockers
The cooling performance of the chassis can be improved by installing 
optional slot blockers. Refer to 
ni.com
 for more details.
Rack Mounting
Rack mount applications require the optional rack mount kits available 
from National Instruments. Refer to the instructions supplied with the rack 
mount kits to install your NI PXIe-1065 chassis in an instrument rack. 
Refer to Figure A-3, 
Note
You may want to remove the feet from the NI PXIe-1065 chassis when rack 
mounting. To do so, remove the screws holding the feet in place. 
Connecting Safety Ground
Caution
The NI PXIe-1065 chassis are designed with a three-position NEMA 5-15 style 
plug for the U.S. that connects the ground line to the chassis ground. To minimize shock 
hazard, make sure the electrical power outlet you use to power the chassis has an 
appropriate earth safety ground.