Chapter 2
Installation and Configuration
© National Instruments Corporation
2-5
The power switch allows you to power on the chassis or place it in standby 
mode. Push the power switch to the On (Left) position (if not already on). 
Observe that the fan becomes operational and the power switch LED is a 
steady green. 
Installing a PXI Controller
This section contains general instructions for installing a PXI controller in 
the PXI-1056 chassis. Refer to your PXI controller user manual for specific 
instructions and cautions. Complete the following steps to install a 
controller.
1.
Plug in the PXI chassis before installing the controller. The power cord 
grounds the chassis and protects it from electrical damage while you 
install the controller. Make sure the chassis power switch is in the 
Off (Standby) position.
Caution
To protect both yourself and the chassis from electrical hazards, leave the chassis 
off until you finish installing the controller.
2.
Install the controller into the system controller slot (slot 1, indicated by 
the red card guides) by first placing the controller edges into the front 
controller guides (right and left). Slide the controller to the rear of the 
chassis (making sure that the injector/ejector handle is pushed down as 
shown in Figure 2-2).