3M™ Rapid Detection Reader 
Running Quality Control 
12 
Running Quality Control 
Two types of quality control test records are stored by the 3M Rapid Detection Reader: 
• 
LQC (Liquid Quality Control) - external surrogate sample test results. 
• 
IQC (Internal Quality Control) - self-diagnostic test results including a check of the power 
supply voltage, system memory, cartridge transport system, cartridge barcode sensor, LED 
function, and incubator function for heated assays. 
Running LQC 
1.  From the Main Menu screen (Figure 4) touch LQC.  The LQC – 
Assay List screen will open (Figure 9).  This screen will display 
the entire list of assays sorted in ascending order by the Time to 
Expiry before assay lockout as displayed in the LQC Expiration 
column.  If LQC is due, this column will display “Run x of y levels” 
where x is the number of levels required and y is the number of 
defined LQC levels for that assay.  Both the number of levels 
required and the available defined levels are set using the RCU. 
The LQC status column shows the requirements for any LQC 
levels that have failed.  If any LQC fails, that specific control level 
must be run successfully prior to proceeding with a patient’s test.  
If the maximum number of Failures before LQC Lockout is 
reached, this column will display “LQC Lockout”. 
NOTE:  The assay list is defined in RCU.  Any new assays added 
in the RCU or through the CM will default to LQC disabled and 
will be listed with N/A in the LQC Expiration column.  Refer to the 
RCU Help Manual for further instruction. 
2.  Select the assay for which LQC will be performed.  Touch Run 
LQC.  If prompted, enter Operator ID and/or Password using 
either the touch screen or the barcode scanner.  To accept the 
displayed information touch Accept.  
NOTE:  Both the Operator ID and Password are case-sensitive. 
Figure 9.  LQC - Assay List 
 
 
Figure 10.  Control ID Entry 
 
 
 
3.  When prompted (Figure 10), select the appropriate Control ID using the drop down menu.  Touch 
Accept. 
NOTE:  If LQC has failed for the assay selected, the number of LQC attempts allowed prior to 
assay lockout is displayed. 
4.  When prompted to Add Control & Insert the Test Cartridge, the LED on a free port of the TM will 
flash blue and the corresponding port icon 
 on the display will illuminate blue (Refer to Test 
Kit and Quality Control Package Inserts for detailed instructions). 
5.  Prepare Test Cartridge as per Test Kit Package Insert and insert Test Cartridge into the flashing 
port. 
If the cartridge is not inserted within 5 minutes of the Add Control & Insert prompt, the system 
will time-out and return to the Main Menu. 
NOTE:  The Test Cartridge must be inserted into the TM within ~30 seconds of sample 
application.  Otherwise, FAIL will be displayed in the port icon and the assay will be aborted.