3M™ Rapid Detection Reader 
Running Quality Control 
13 
Do not use excessive force when inserting the Test Cartridge.  The TM will guide the Cartridge 
into place.  
Once the Cartridge Inserted message clears, the previous screen is displayed and the active port 
icon 
 shows a timer that counts up until sample flow is detected, then counts down until the 
assay is complete. 
NOTE:  Touching Run Assay or Run LQC will start another assay/LQC and the test sequence 
will begin again using a different available port.  The assay progress can be checked by touching 
the port icon.  Touch Exit to return to the previous screen. 
6.  When an assay is completed, the corresponding port icon displays 
 and the Test Cartridge 
is ejected.  Touch the icon to view the test details. 
7.  Remove the Test Cartridge and discard according to local hazardous waste policy. 
 
LQC Icons 
Icon 
Meaning 
 
LQC will be due shortly.  Appears when the LQC timer for an assay has less than the 
user defined % of time remaining until lockout. 
 
LQC has expired for one or more assays.  If icon is 
displayed, the Assay Status Warnings screen 
(Figure 11) will display upon touching the Run Assay 
button.  Touch Continue if the desired assay is not 
listed. 
A limited number of overrides can be configured via 
the RCU.  If overrides are available, touch Continue 
to override lock out (i.e., perform a patient test after 
LQC is due).  The decision to use an Override will be 
confirmed (Figure 12).  Once all overrides have been 
used, LQC must be performed prior to any further 
patient testing.  If an assay with no overrides is 
inserted into the TM, an LQC Overdue, Assay 
Locked 
warning will display and the assay will be 
aborted.  
Figure 11.  Assay Status Warnings 
 
 
Figure 12.  Override Confirmation 
 
 
NOTE:  These warnings are display only and are not selectable.  To determine the assay(s) for 
which LQC is due, view the LQC – Assay List (Figure 9). 
NOTE:  If there is more than one level of LQC defined for an assay, and LQC has expired for this 
assay, the LQC timer will be reset to the time at which the earliest LQC level passes in the set.  
e.g.:  Two levels of QC are required, Level 2 passes LQC but Level 1 fails; once Level 1 is 
repeated and passes, the "Time to Expiry" column in the LQC menu will display the count down 
from the time when Level 2 passed.