3M™ Rapid Detection Reader 
Managing Settings 
20 
Settings: View Tabs 
Tab 
Screen 
LQC continued 
Max Failures Before Lockout 
Purpose:  Define the number of LQC failures allowed before the 
assay is Locked Out when LQC Timer is set in RCU.  LQC 
Failure count is reset upon successful completion of LQC.  Can 
be set from 1-9.   
Default:  9. 
Control ID 
Purpose:  Define control IDs.   
Default:  Min length 1, Max length 18. 
 
Timer  
The ability to edit the settings of the TIMER tab varies by item 
and is defined below.  
Timer Intervals  
Menu Idle Timeout 
Edit on CM.  Cannot edit through RCU. 
Purpose:  Define the amount of idle time before the 3M Rapid 
Detection Reader returns to the Home screen.  
Range:  1 -10 minutes.  
Default:  5 minutes. 
Update/Upload Start Date/Time 
Edit through RCU only.  
The date and time used to calculate the start of the following 
intervals. 
Update Interval 
Edit through RCU only. 
Purpose:  Interval at which the Reader checks for changes to 
settings in a networked environment.  Disable by entering 0.  
Range: 0 – 1440 minutes. 
Default: 1.  
Upload Results Interval 
Edit through RCU only. 
Purpose:  Interval at which the Reader sends changes or 
additions to the result and event logs in a networked environment.  
Disable by entering 0.  
Range:  0 - 1440 minutes.  
Default:  0 (Disabled).  
Figure 23.  Settings: View Timer 
Tab