3M™ Rapid Detection Reader 
Managing Settings 
18 
Managing Settings 
The 3M Rapid Detection Reader can be operated immediately after set-up using the pre-
programmed default settings.  Each Reader also has customizable features which can be 
configured to enhance its functionality and data management capability.  Customizable features 
include LQC timer, Operator ID and password, and data transfer options (USB, print, network).  
Some of these features can be configured in the CM Settings menu.  In the default configuration, 
these settings can only be viewed or changed by an operator with Supervisor access.  Most 
feature configurations are set using the Reader Configuration Utility (RCU) by a Full Access User.  
For instruction on configuring settings on the CM, refer to Edit Settings and Edit Operators.  For 
instructions on configuring settings using the RCU, refer to Editing Settings Using the Reader 
Configuration Utility (RCU)
. 
 
Feature (See Detailed 
Descriptions Below) 
Full Access User 
Edit on RCU 
Supervisor 
Edit on CM 
Assay defaults  
√ 
 
LQC 
√ 
 
Update / Upload Features 
√ 
 
Printer Settings 
√ 
 
Login Requirements 
√ 
√ (limited) 
File Settings 
√ 
√ 
Menu Idle / Timeout 
 
√ 
Date / Time (without Network 
Time Protocol – NTP) 
NTP only 
√ 
Language 
 
√ 
Auto IQC Interval 
√ 
√ 
Network Transfer Timers 
 
√ 
Sound 
 
√ 
 
Accessing the Settings Menus 
From the Main Menu (Figure 4), touch Settings.  Access to the 
Settings menu is Operator ID and Password protected.  The default 
login for access to the Settings menus is:  
Operator ID:  SUPERVISOR 
PASSWORD:  PASSWORD 
Both entries are case-sensitive.  To provide additional security, it is 
highly recommended that this login be changed.  
Figure 20.  Settings Menu 
 
 
NOTE:  The default privilege access to the Settings menu is Supervisors only (Figure 25).  The 
privilege access to this menu can only be changed via the RCU.  In the default setting, any operator 
defined in the Operator List as Supervisor [Operator Type] is able to access the Settings menu.  Refer 
to Managing Operators for additional information.  
Touch View Settings from the Settings menu (Figure 20).  This will open the screen shown in Figure 
21.  Settings in the View Settings screen cannot be edited.