3M™ Rapid Detection Reader 
Troubleshooting 
43 
Settings Messages 
Message 
Cause 
Action 
User Expired, Try Again 
User certification has expired or 
the details were entered 
incorrectly. 
Re-enter Operator ID or contact 
Supervisor. 
 
LQC Related Messages 
Message 
Cause 
Action 
LQC Lockout. Ejecting. 
 
LQC is overdue for this assay 
and no overrides are available.  
LQC must be run and passed, 
or a Full Access User can reset 
the LQC timer and/or overrides 
for this assay in Reader 
Settings on the RCU. 
Invalid Control. Ejecting. 
Control ID does not exist for this 
assay or has been entered 
incorrectly. 
Repeat LQC with correct 
Control ID or contact 
Supervisor. 
LQC Overdue, Assays Locked 
 
LQC has expired and all 
available overrides have been 
used.  Tests cannot be run until 
the required LQC has been run 
and passed. 
Run the required LQC. 
 
LQC will be due shortly.  
Appears when the LQC timer for 
an assay has less than the user 
defined % of time remaining until 
lockout. 
Run the required LQC. 
 
LQC has expired for one or more 
assays.  If icon is displayed, the 
Assay Status Warnings screen 
(Figure 11) will display upon 
touching the Run Assay button.   
 
Run the required LQC. 
A limited number of overrides 
can be configured via the RCU.  
If overrides are available, touch 
Continue to override lock out 
(i.e., perform a patient test after 
LQC is due).  The decision to 
use an Override will be 
confirmed (Figure 12).  Once all 
overrides have been used, LQC 
must be performed prior to any 
further patient testing.  If an 
assay with no overrides is 
inserted into the TM, an LQC 
Overdue, Assay Locked 
warning will display and the 
assay will be aborted.