Справочник Пользователя для 3M #3968

Скачать
Страница из 58
3M™ Rapid Detection Reader 
Troubleshooting 
42 
Settings Messages 
Message 
Cause 
Action 
Access denied. Try again. 
 
The user does not have the 
login permissions necessary to 
access selected function. 
Confirm correct login details and 
try again. 
Assay List Size Exceeded. 
 
The list of assays permitted is 
full.  No further additions can be 
made. 
Delete existing assays from the 
assay list (Refer to RCU online 
help).  The maximum number of 
assay entries allowed is 200. 
Assay Locked, Aborting. 
 
The assay is not permitted or 
has been locked.  May occur if 
LQC has failed or assay is not in 
the assay list. 
Contact Supervisor or RCU Full 
Access User for more 
information. 
Database Almost Full, Backup 
Data 
Displayed when the3M Rapid 
Detection Reader is not 
connected to a network and 
either the patient or LQC result 
databases has less than 20 
spaces left.  This message will 
continue to be displayed until 
the database is cleared.  
Transfer data to USB device 
and delete the database. 
Default Restore Failed. 
The action to restore defaults 
has failed.  
Try this process again.  If 
problem persists contact 
Technical Support. 
ERROR: Invalid Code. Try 
Again. 
The override code entered is 
invalid. 
Contact Technical Support, then 
enter the override code 
provided.  
ERROR: Numbers Only.  
Only numbers are permitted in 
the override code field. 
Enter the number provided by 
Technical Support. 
Invalid Password, Try Again. 
The password was incorrectly 
entered. 
Retype the password and try 
again. 
Invalid User, Try Again. 
The user is not recognized or 
has been entered incorrectly. 
Retype user details and try 
again. 
Locked, Contact Supervisor. 
The 3M Rapid Detection Reader 
has been locked via the RCU.  
Contact the RCU Full Access 
User to unlock using the Reader 
Settings screen. 
Empty Search. 
The 3M Rapid Detection Reader 
cannot find the record 
requested.  It may have been 
deleted or details entered 
incorrectly. 
Re-enter search data and try 
again.