Chapter 2
Function Reference — SCXI_Cal_Constants
NI-DAQ FRM for PC Compatibles
2-328
©
 National Instruments Corporation
use the endpoints of your expected input range as scaled1 and scaled2. Then you would be 
specifically calibrating your expected input voltage range.
If you are using an SCXI-1100, SCXI-1122, SCXI-1126, or SCXI-1141, you can perform a 
one-point calibration to determine the binary offset; you can do this easily without external 
hookups using the 
SCXI_Calibrate_Setup
 function to internally ground the amplifier. 
Use the procedure above, skipping steps 6 and 7, and using opCode = 1 for the 
SCXI_Cal_Constants
 function.
If you are storing calibration constants in the SCXI-1122, SCXI-1126, or SCXI-1141 
EEPROM, your binary offset and gain adjust factors must not exceed the ranges given in 
the respective module user manuals. The constant format in the EEPROM does not allow 
for larger constants. If your constants exceed these specifications, the function returns 
badExtRefError. If this error occurs, you should make sure your SCXIgainDAQgain, and 
TBgain values are the actual settings you used to measure the volt/binary pairs, and you might 
want to recalibrate your DAQ device, if applicable.
Analog Output Calibration
When you call 
SCXI_AO_Write
 to output a voltage or current to your SCXI-1124 module, 
NI-DAQ uses the calibration constants loaded for the given module, channel, and output range 
to scale the voltage or current value to the appropriate binary value to write to the output 
channel. By default, NI-DAQ will load calibration constants into memory for the SCXI-1124 
from the module EEPROM load area (see the 
 section for more 
information).
You can recalibrate your SCXI-1124 module to create your own calibration constants using 
the following procedure:
1.
Use the 
SCXI_AO_Write
 function with opCode = 1. If you are calibrating a voltage 
output range, pass the parameter binaryData = 0. If you are calibrating the 0 to 20 mA 
current output range (rangeCode = 6), pass the parameter binaryData = 255.
2.
Measure the output voltage or current at the output channel with a voltmeter. This is your 
first volt/binary pair: binary1 = 0 or 255 and volt1 is the voltage or current you measured 
at the output.
3.
Use the 
SCXI_AO_Write
 function with opCode = 1 to write the binaryData = 4,095 to 
the output DAC.
4.
Measure the output voltage or current at the output channel. This is your second 
volt/binary pair: binary2 = 4,095 and volt2 is the voltage or current you measured at the 
output.
5.
Call 
SCXI_Cal_Constants
 with your voltage/binary pairs and opCode = 2. You can 
save the constants on the module EEPROM (calibrationArea = 1 or 3). Refer to the 
following 
 section for information about constants in the 
EEPROM. It is best to use calibrationArea = 3 (user EEPROM area) as you are