Chapter 2
Function Reference — SCXI_Calibrate_Setup
NI-DAQ FRM for PC Compatibles
2-330
©
 National Instruments Corporation
SCXI_Calibrate_Setup
Format
status = SCXI_Calibrate_Setup (SCXIchassisID, moduleSlot, calOp)
Purpose
Used to ground the amplifier inputs of an SCXI-1100, SCXI-1122, or SCXI-1141 so that you 
can determine the amplifier offset. You can also use this function to switch a shunt resistor 
across your bridge circuit to test the circuit. Shunt calibration is supported for the SCXI-1122 
or SCXI-1121 modules with the SCXI-1321 terminal block.
Parameters
Input
Parameter Discussion
calOp indicates the calibration mode you want.
0:
Disable calibration.
1:
Connect the positive and negative inputs of the SCXI-1100, SCXI-1122, 
or SCXI-1141 amplifier together and to analog reference.
2:
Switch the shunt resistors across the bridge circuit on the SCXI-1121 
(Revision C or later) or SCXI-1122.
Using This Function
The zero offset of the SCXI-1100, SCXI-1122, or SCXI-1141 amplifiers varies with the 
module gain. When you know the offset at a specific gain setting, you can add that offset to 
any readings acquired at that gain. In general, the procedure for determining the offset at a 
particular gain is as follows:
1.
SCXI_Single_Chan_Setup
—Enable the module output, route the module output 
on the SCXIbus if necessary, and resolve any SCXIbus contention if necessary. For 
the SCXI-1100 and SCXI-1122, the module channel you specify is irrelevant.
2.
SCXI_Set_Gain
—Set the module gain to the setting that you will use in your 
application.
3.
SCXI_Calibrate_Setup
—Ground the amplifier inputs.
Name
Type
Description
SCXIchassisID
i16
logical ID assigned to the SCXI chassis
moduleSlot
i16
chassis slot number
calOp
i16
calibration mode