Chapter 2
Function Reference — SCXI_Track_Hold_Setup
NI-DAQ FRM for PC Compatibles
2-370
©
 National Instruments Corporation
DAQdeviceNumber is the device number of the DAQ device in the PC that will be used to 
acquire the data. If the DAQdeviceNumber specified is a Lab and 1200 Series devices, 
DAQCard-700, or LPM device, inputMode 2 is not supported. If you are using the 
SCXI-1200 to acquire the data, use the logical device number you assigned to the SCXI-1200 
in the configuration utility.
Using This Function
For single channel operations (inputMode = 1) the module is level-sensitive to the source 
signal; that is, when the source signal is low the module is in track mode, and when the 
source signal is high the module is in hold mode. If source = 0, you can use calls to 
SCXI_Track_Hold_Control
 function to put the module into track or hold mode by 
toggling the output of the appropriate counter on the DAQ device. If the SCXI-1140 you want 
to read is not cabled to the DAQ device, you will have to configure the SCXI-1140 module 
that is cabled to the DAQ device to send the counter output on the SCXIbus to the module you 
want. Then the 
SCXI_Track_Hold_Control
 call can put the module you want into track or 
hold mode. The 
SCXI_Track_Hold_Setup
 parameters for each module would be:
For the SCXI-1140 that is cabled to the DAQ device as follows:
inputMode = 1
source = 0
send = 2
For the SCXI-1140 module to be read:
inputMode = 1
source = 2
send = 0
Using an external source (source = 1) for single channel operations is not normally useful 
because NI-DAQ has no way of determining when the module has gone into hold mode and 
it is appropriate to read the channels.
(MIO, Lab-PC-1200, Lab-PC-1200AI, Lab-PC+, PCI-1200, SCXI-1200, and 
DAQCard-1200 only) For interval channel scanning operations (inputMode = 2) NI-DAQ 
configures the module to go into hold mode on the rising edge of the source signal. If 
source = 0, that will happen when counter 2 on the Am9513-based MIO devices, a dedicated 
DAQ-STC counter on E Series devices, or counter B1 on the Lab-PC-1200, Lab-PC-1200AI, 
Lab-PC+, PCI-1200, SCXI-1200, or DAQCard-1200 pulses at the beginning of each scan 
interval; if source = 1, that will happen on the rising edge of the external signal connected to 
HOLDTRIG on the module front connector. In the latter case, you should configure the DAQ 
device for external scan interval timing (using the 
DAQ_Config
 function) so that the external 
signal will trigger each scan. To scan more than one SCXI-1140, you can send the source 
signal from the module that is receiving it (either from the counter or from HOLDTRIG) to 
the other modules over the SCXIbus. Notice that the module that is cabled to the device can