Troubleshooting
En
glish
Multipatient use
WARNING
An antibacterial filter is mandatory if the device is used on multiple patients.
In a mulitpatient use environment, you must perform the following before the device is 
provided to a new patient:
Servicing
CAUTION
Inspection and repair should only be performed by an authorised agent. Under no 
circumstances should you attempt to open, service or repair the device yourself.
This product should be inspected by an authorised ResMed service centre five years from 
the date of manufacture, except for the internal battery which must be replaced every two 
years from the manufacturing date of the device. Prior to this, the device is intended to 
provide safe and reliable operation provided that it is operated and maintained in 
accordance with the instructions provided by ResMed. Applicable ResMed warranty 
details are provided with the device at the time of original supply. Of course, as with all 
electrical devices, if any irregularity becomes apparent, you should exercise caution and 
have the device inspected by an authorised ResMed service centre.
Troubleshooting
If there is a problem, try the following suggestions. If a problem cannot be solved, contact 
ResMed. 
Alarm troubleshooting
The most common reason for an alarm to sound is because the system has not been 
properly assembled. Check that the air tubing has been properly attached to the device and 
patient interface (and humidifier if used).
Notes: 
• The alarm log and alarm settings are maintained when the device is powered down and 
in the event of a power loss.
• If multiple alarms are active simultaneously, the alarm with the highest priority will be 
displayed first.
• If an alarm activates repeatedly, discontinue use and return the device for servicing.
Air filter and the 
antibacterial filter
Replace.
Mask
Reprocess; Cleaning, disinfection and sterilisation 
instructions are available from the ResMed website, 
www.resmed.com/masks/sterilization. If you do not have 
Internet access, please contact your ResMed representative.
Air tubing
Replace the air tubing. Alternatively, consult the air tubing 
instructions for cleaning and disinfection information.
Device
Disinfect the Stellar as follows:
1. Apply undiluted mikrozid
®
 AF or CaviCide
®
 liquid to a 
clean non-dyed disposable cloth.
2. Wipe all surfaces of the device, including the air outlet; 
avoid liquid entering any openings in the device.
3. Leave the disinfectant to work for five minutes.
4. Wipe residual disinfectant from the device with a clean, 
dry non-dyed disposable cloth.
Humidifier
As instructions for humidifiers vary, see the user guide for 
the humidifier in use.