Troubleshooting
En
glish
LCD: Circuit disconnected!
The air tubing is not connected properly
to the humidifier or the device.
to the humidifier or the device.
1. Check that the air tubing is connected
properly to the humidifier or the
device.
device.
2. If the alarm is not cleared, stop
treatment.
3. Re-start treatment.
LCD: Low Minute Ventilation!
Minute ventilation level has dropped
below the alarm setting level.
below the alarm setting level.
Contact your clinician.
LCD: Low Respiratory Rate!, High Respiratory Rate!
The respiratory rate level has dropped
below or has exceeded the alarm setting
level.
below or has exceeded the alarm setting
level.
Contact your clinician.
LCD: High Leak!
High mask leak for more than 20
seconds.
seconds.
• Adjust the mask to minimise leak (see
• Check the air circuit integrity and
reconnect.
• If the problem persists contact your
clinician.
LCD: Non-Vented Mask!
• Connection of a non-vented mask.
• Mask vents may be blocked.
• ResMed leak port is missing or vent is
• Mask vents may be blocked.
• ResMed leak port is missing or vent is
blocked.
• Ensure the mask has vents.
• Ensure the mask vents are not
• Ensure the mask vents are not
blocked.
• Ensure the leak port is installed and
that the vent is not blocked.
• Ensure oxygen (if in use) has only been
connected at the rear of the device.
• If the problem persists contact your
clinician.
LCD: Apnoea!
The device detects an apnoea that has
exceeded the pre-set alarm level.
exceeded the pre-set alarm level.
• Breathe normally to disable the alarm.
• If the problem persists contact your
• If the problem persists contact your
clinician.
LCD: Internal Battery Low!
The internal battery capacity is below
30%.
30%.
Connect the device to mains power.
LCD: Low SpO
2
!
SpO
2
has dropped below pre-set alarm
level.
• Check the attachment of the sensor.
• If the problem persists contact your
• If the problem persists contact your
clinician.
LCD: SpO
2
finger sensor failure!
The finger sensor is not connected
properly or delivers faulty values.
properly or delivers faulty values.
Check if the finger sensor is attached
properly to the finger and connection to
the pulse oximeter.
properly to the finger and connection to
the pulse oximeter.
LCD: Xpod oximeter disconnected!
The pulse oximeter is disconnected.
Check if the pulse oximeter is connected
properly to the device.
properly to the device.
Problem / possible cause
Action