Troubleshooting
En
glish
LCD: Circuit disconnected!
The air tubing is not connected properly 
to the humidifier or the device. 
1. Check that the air tubing is connected 
properly to the humidifier or the 
device.
2. If the alarm is not cleared, stop 
treatment.
3. Re-start treatment. 
LCD: Low Minute Ventilation!
Minute ventilation level has dropped 
below the alarm setting level.
Contact your clinician.
LCD: Low Respiratory Rate!, High Respiratory Rate!
The respiratory rate level has dropped 
below or has exceeded the alarm setting 
level.
Contact your clinician.
LCD: High Leak!
High mask leak for more than 20 
seconds.
• Adjust the mask to minimise leak (see 
• Check the air circuit integrity and 
reconnect.
• If the problem persists contact your 
clinician.
LCD: Non-Vented Mask!
• Connection of a non-vented mask.
• Mask vents may be blocked.
• ResMed leak port is missing or vent is 
blocked.
• Ensure the mask has vents.
• Ensure the mask vents are not 
blocked.
• Ensure the leak port is installed and 
that the vent is not blocked.
• Ensure oxygen (if in use) has only been 
connected at the rear of the device.
• If the problem persists contact your 
clinician.
LCD: Apnoea!
The device detects an apnoea that has 
exceeded the pre-set alarm level. 
• Breathe normally to disable the alarm.
• If the problem persists contact your 
clinician.
LCD: Internal Battery Low!
The internal battery capacity is below 
30%.
Connect the device to mains power.
LCD: Low SpO
2
!
SpO
2
 has dropped below pre-set alarm 
level.
• Check the attachment of the sensor.
• If the problem persists contact your 
clinician.
LCD: SpO
2
 finger sensor failure!
The finger sensor is not connected 
properly or delivers faulty values.
Check if the finger sensor is attached 
properly to the finger and connection to 
the pulse oximeter.
LCD: Xpod oximeter disconnected!
The pulse oximeter is disconnected. 
Check if the pulse oximeter is connected 
properly to the device.