Notes: 
• The manufacturer reserves the right to change these specifications without notice.
• Pressure may be displayed in cm H
2
O or hPa.
Guidance and manufacturer’s declaration – electromagnetic 
emissions and immunity
Medical Electrical Equipment needs special precautions regarding EMC and needs to be 
installed and put into service according to EMC information provided in this document.
IEC 60601-1 
classifications
• Class II (Clause 3.14—double insulation). This 
adherence means the need for an protective earthing 
(ie, an earthed plug) is not necessary. 
• Type BF
• Continuous operation
Air travel requirements
Medical-Portable Electronic Devices (M-PED) that meet 
the Federal Aviation Administration (FAA) requirements 
of RTCA/DO-160 can be used during all phases of air 
travel without further testing or approval by the airline 
operator. ResMed confirms that the Stellar meets RTCA/
DO-160 requirements.
 
This device is not suitable for use in the presence of a flammable anesthetic mixture.  
Guidance and manufacturer’s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the 
device should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR11 
Group 1
The device uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not 
likely to cause any interference in nearby electronic 
equipment.
RF emissions CISPR 11 
Class B
The device is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage network that 
supplies buildings used for domestic purposes.
Harmonic Emissions 
IEC 61000-3-2
Class A
Voltage Fluctuations/Flicker 
Emissions IEC 61000-3-3
Complies
Warnings: The device should not be used adjacent to or stacked with other equipment. 
If adjacent or stacked use is necessary, the device should be observed to verify normal operation in the 
configuration in which it will be used.
The use of accessories (eg, humidifiers) other than those specified in this manual is not recommended. They may 
result in increased emissions or decreased immunity of the device.
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the 
device should assure that it is used in such an environment.
Immunity test
IEC60601-1-2 test 
level
Compliance level
Electromagnetic environment –guidance
Electrostatic 
discharge (ESD) 
IEC 61000-4-2
±6 kV contact 
 
±8 kV air
±6 kV contact 
 
±8 kV air
Floors should be wood, concrete or ceramic tile. 
If floors are covered with synthetic material, the 
relative humidity should be at least 30%.
Electrical fast 
 
transient/burst
IEC 61000-4-4
±2 kV for power 
supply lines
±1 kV for input/output 
lines
±2 kV 
 
±1 kV
Mains power quality should be that of a typical 
commercial or hospital environment.
Surge IEC 
61000-4-5
±1 kV differential 
mode
±2 kV common mode
±1 kV differential 
mode 
±2 kV common 
mode
Mains power quality should be that of a typical 
commercial or hospital environment.