Справочник Пользователя для ResMed 2011-09

Скачать
Страница из 38
29
Technical specifications
En
glish
  
Voltage dips, 
short 
interruptions  
and voltage 
variations on 
power supply 
input lines 
IEC 61000-4-11
<5% Ut (>95% dip in 
Ut) for 0.5 cycle 
40% Ut (60% dip in 
Ut) for 5 cycles 
70% Ut (30% dip in 
Ut) for 25 cycles 
<5% Ut (>95% dip in 
Ut) for 5 sec 
< 12 V (>95% dip in 
240V) for 0.5 cycle
96 V (60% dip in 
240 V) for 5 cycles 
168 V (30% dip in 
240 V) for 25 cycles 
<12 V (>95% dip in 
240 V) for 5 sec
Mains power quality should be that of a typical 
commercial or hospital environment.
If the user of the device requires continued 
operation during power mains interruptions, it is 
recommended that the device be powered from 
an uninterruptible power source.
Power 
frequency 
 
(50/60 Hz) 
magnetic field
IEC 61000-4-8
3 A/m 
3 A/m 
Power frequency magnetic fields should be at 
levels characteristic of a typical location in a 
typical commercial or hospital environment.
Portable and mobile RF communications 
equipment should be used no closer to any part 
of the device, including cables, than the 
recommended separation distance calculated 
from the equation applicable to the frequency of 
the transmitter.
Recommended separation distance: 
Conducted RF 
 
IEC 61000-4-6
3 Vrms 
 
150 kHz to 80 MHz
3 Vrms
 d = 1.17 
√P
Radiated RF
 
IEC 61000-4-3
10 V/m
80 MHz to 2.5 GHz
3 V/m
d = 0.35 
√P 80 MHz to 800 MHz
d = 0.70 
√P 800 MHz to 2.5 GHz
where P is the maximum output power rating of 
the transmitter in watts (W) according to the 
transmitter manufacturer and d is the 
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey,
a
 
should be less than the compliance level in each 
frequency range.
b
 
 
Interference may occur in the vicinity of 
equipment marked with the following symbol: 
 
NOTE 1: Ut is the AC mains voltage prior to application of the test level.
NOTE 2: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 
from structures, objects and people.
a
 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the  
measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device 
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, 
such as reorienting or relocating the device.
b
 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the 
device
The device is intended for use in an environment in which radiated RF disturbances are controlled. The customer or 
the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between 
portable and mobile RF communications equipment (transmitters) and the device as recommended below, according 
to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter (m)
Rated maximum output 
power of transmitter (W)
150 kHz to 80 MHz 
d = 1.17 
√P
80 MHz to 800 MHz
d = 0.35 
√P
800 MHz to 2.5 GHz
d = 0.7 
√P
0.01
0.17
0.04
0.07
0.1
0.37
0.11
0.22
1
1.17
0.35
0.7