Справочник Пользователя для Sierra 840

Скачать
Страница из 74
Chapter 1 Introduction 
Series 830/840/860 Instruction Manual
 
1-4 
IM-83/84/86-H
 
The Series 800 Flow Sensing Principle 
The operating principle of Sierra’s Side-Trak and Auto-Trak flow 
meters and controllers is based on heat transfer and the first law of 
thermodynamics. During operation process gas enters the instru-
ment’s flow body and divides into two flow paths, one through the 
sensor tube, the other through the laminar flow element bypass. 
The laminar flow element bypass generates a pressure drop, P
1
–P
2
forcing a small fraction of the total flow to pass through the sensor 
tube (m
1
) which is then monitored. 
 
 
Figure 1-1. Flow Paths through the Transducer 
 
Two resistance temperature detector (RTD) coils around the sensor 
tube direct a constant amount of heat (H) into the gas stream. In ac-
tual operation, the gas mass flow carries heat from the upstream 
coil to the downstream coil. The resulting temperature difference 
(∆T) is detected by the RTD coils. 
 
 
Figure 1-2. Flow Measuring Principle