Series 830/840/860 Instruction Manual 
Chapter 2 Installation
 
IM-83/84/86-H 
2-1
 
Chapter 2  Installation 
Quick-Installation Instructions For  
830/840 L, M and H Meters And Controllers
 
 
1.  NEVER TEST FOR LEAKS WITH LIQUID LEAK DETECTOR.  If 
liquid seeps into the electronics or the sensor compartment below, 
the  instrument  may  be  damaged.    Use  a  pressure-decay  test  in-
stead. 
2.  Install an appropriate in-line filter upstream if the gas contains 
any  particulate  matter  or  condensed  moisture.  (A  15-micron  par-
ticulate  filter  for  full-scale  flows  up  to  30  liters/minute  is  recom-
mended, 30 micron for flows above 30). 
3.  Mount with a horizontal gas-flow.  This orientation is preferable 
unless the factory calibration was specifically performed for a ver-
tical  flow.    Consult  your  calibration  certificate.    (Horizontal  flow  is 
assumed unless vertical flow upward or downward is specified). 
4.  DO  NOT  APPLY  POWER  TO  THE  OUTPUT  LOOP  on  units 
equipped and calibrated for a 4-20 mA output signal.  This is NOT 
a loop-powered device.  Damage will occur.  
5.  Wire the instrument per the diagrams on the back of this card or 
the Operator’s Manual, Chapter 2.  Controllers require a separate 
ground wire for the Valve Return signal as well as a Setpoint sig-
nal (no wire connected to the setpoint may cause the valve to float 
open).   
6.  Apply the gas listed on the label to the inlet at the recommended 
inlet pressure as listed on the calibration certificate.  On a meter, 
obtain  this  pressure  by  blocking  off  the  flow  downstream.    On  a 
controller,  confirm  that  there  is  no  gas  flowing  through  it  with  a 
zero  setpoint.    (If  there  is  a  flow,  consult  the  Valve  Adjustment 
Procedure in the manual). 
7.  Apply power and verify or adjust the zero setting after allowing 
a 15-minute warm-up period.  The zero adjustment potentiometer 
is  accessed  through  the  upper  hole  in  the  side  of  the  electronics 
cover. (See the Operator’s Manual, Chapter 3). 
8.  DO  NOT  LEAVE  A  SETPOINT  APPLIED  TO  A  CONTROLLER 
WHEN  NO  GAS  IS  AVAILABLE  TO  THE  INLET  FITTING.    The 
control circuit will apply the maximum voltage to the valve coil re-
sulting  in  eventual  overheating.    Damage  may  occur.    (Instead, 
consult the Operator’s Manual for use of the “Valve Off” feature). 
9. 
An  ANNUAL  factory  evaluation  and  calibration  is  recom-
mended.
 
 
 Email Technical Support:  Service@sierrainstruments.com   
Website Self Service:  www.sierrainstruments.com