Chapter 2 Installation 
Series 830/840/860 Instruction Manual
 
2-2 
IM-83/84/86-H
 
Installation Overview 
Side-Trak
 and Auto-Trak
 transducers are supplied with either 
compression, VCO, VCR or NPT process connections. To ensure a 
successful installation, inlet and outlet tubing should be in a clean 
state prior to plumbing the transducer into the system. The ship-
ping caps covering the inlet/outlet fittings should not be removed 
until immediately prior to installation. 
 
Before installing the transducer, verify the following: 
 
1.  Make sure the installation site meets the specific operating pa-
rameters recorded on the transducer’s nameplate. Each trans-
ducer is factory-configured for a specific gas and flow range, 
pressure differential, temperature range and mounting position. 
 
2.  Do not locate the transducer in areas subject to sudden tem-
perature changes, moisture, drafts or near equipment radiating 
significant amounts of heat. Make sure to allow adequate space 
for cable connectors and wiring. 
 
3.  Make sure the location meets the minimum number of recom-
mended pipe diameters upstream and downstream of any NPT 
or 1/2-inch process connection. A minimum of two inches is 
always recommended. 
 
4.  If the gas contains any particulate matter, install an in-line fil-
ter prior to the transducer. Recommended filter size: 15 mi-
cron for flows of 10 to 30 slpm, 30 micron for above 30 slpm. 
 
5.  If the transducer is mounted in any position other than horizontal 
and it was not calibrated specifically for that position, contact Si-
erra’s Customer Service Department for instructions. 
 
6.  For meters, if a potential over-flow condition exists, insert a 
valve or critical orifice in the line to limit flow to approxi-
mately 25 percent above the full scale range of the meter. 
 
7.  For controllers, use a properly-sized pressure controller and 
verify the controller orifice size is the smallest in the system. 
There can be no restrictions (such as valves, tubing or piping 
internal diameters, reducers, etc.) upstream or downstream of 
the controller less than the valve orifice diameter. 
 
8.  Confirm that the transducer o-ring material is compatible with 
the gas to be measured.